Microscope optics V2

This guide shows the current specification for all the lensing in the microscope.

It lists all the lenses used in the microscope as well as the exact positioning for each of the lenses to achieve the desired projection and camera magnification.

For each lens choice it estimated the best value using screenshots generated by running the Optics App with test values.

Specs

Here’s a list of all the lenses used in the microscope. All distances are from the center of the dichroic cube. The axis from projector to tissue goes negative to positive. The axis from tissue to camera is also negative to positive.

  1. Projector. PROPixx projector. Projected image is at 1.22 meters from projector (@67cm). @-55cm.

  2. L1. AC508-075-A-ML. f=75mm – 2”. @-51cm.

  3. L2. AC508-080-A-ML. f=80mm – 2”. @-49cm.

  4. ND filter ring.

  5. L3. AC508-200-A-ML. f=200mm – 2”. @-27cm.

  6. Dichroic.

  7. L4. Objective lens @13cm (center). There are two objectives to choose from, a 4x (the default - all computations below use this) and a 2x when we need larger area.

    • Nikon Plan Fluor. F=41.3mm. 4x/0.13, WD17.2.

    • CFI Plan Apochromat Lambda. F=mm. 2x/0.1, WD8.5.

  8. Tissue. @17.1748cm.

  9. L5. AC508-150-A-ML. f=150mm – 2”. @7cm.

  10. Camera. 4070C-USB Thor cam. Sensor plate @21.4077cm.

The measured tissue projected area with the 4x objective is approximately 4.24mm x 2.41mm. For the 2x objective it’s 8.36mm x 4.72mm. Computed magnification from the projector to the tissue is -0.0072 for the 4x objective.

The camera sensor has 2048x2048 pixels and each pixel is 7.4μm x 7.4μm. Sensor area width and height is 15.1552mm. Computed magnification from the tissue to the camera is -3.6403. This results in a computed field of view on the tissue of 4.16mm square.

image0

Figure 1 The microscope

image1

Figure 2 The microscope annotated

Projector to tissue light path

For the projector to tissue path, the axis goes from the projector towards the tissue through the dichroic mirror. The center of the dichroic is the zero point, and the projector is on the negative part of the axis and the tissue is on the positive side.

Static Optical Analysis

Following is the details of all the optical components in the projector path.

L1

L1 must sit close to the projector to get the full projector view and most of the light. The “object” input to the lens is projected by the projector. The “object” is between 1.22m to 10m on the right of the projector, depending on the focal setting of the projector’s lens. Although it’s the right of L1, we use the provided “object” position. We assume it’s set at 1.22m.

image2

Figure 3 The characteristics of L1.

The figure is read as follows, we input the focal length of the lens, the position of the lens and the object (as output by the projector). The output is the estimated image position (on the other side of the lens, if positive relative to the lens), the magnification, and the relative distance between the lens and the image.

L2

L2 sits as close to L1 as possible and together they focus the image just to the right of L2 and at a significant downscaling.

image3

Figure 4 The characteristics of L2.

L3

L3 controls the zoom of the final projection on the tissue as well as the distance from the objective lens to the projection. The goal is for that projection to be at the rated working distance of the objective and to size the projection to the required size.

This lens creates a virtual image at negative infinity. This collimates the light before it reaches the objective.

image4

Figure 5 The characteristics of L3.

L4

L4 is the final objective lens before the tissue. The position of the lens is dictated by the positions of all the previous lenses and the projector, given the objective’s working distance. The projection should be in focus at the working distance.

image5

Figure 6 The characteristics of L4.

Dynamic Optical Analysis

In the following, we explore how the lens values were chosen and how the lens values change the primary variables – the final position of the projected pattern and its magnification.

Projector

The projector is placed as close to the tissue as possible, given the lenses in the path, so that we can capture all the light from the projector and the full view. The projector has inbuilt lenses that diverge the light at a smallish angle as it’s coming out from the projector.

L1-L2

Lens L1 and L2 are the smallest focal length lens at 2” diameter that we can get. Its purpose is to scale down the image as much as possible. The lenses are placed as close to each other as possible. The resultant image should be focused just to the right of L2, as can be seen on a piece of paper.

L3 Dynamic

The purpose of L3 is to provide some further down-scaling of the image. It also collimates the light by forming a virtual image at negative infinity. Light collimation is required for the objective to function optimally.

The exact placement of the lens and its distance from the objective (L4) determines how much down-scaling occurs. It also controls the distance between the objective and the focused image, which ideally will be the working distance of the objective to fully benefit from the objective.

Assuming some placement of L4 (additionally investigated below), we have two variables to control; the location and focal length of L3. The following images investigate these variables.

image7

Figure 7 The final image magnification on the tissue in response to L3 position and focal length. The red square is the selected values for the microscope.

The desired value is 0.005 - 0.01. From the contours we see that increasing the focal length decreases the magnification at this position, with larger “jumps” at larger focal values. However, moving the lens by small values keeps the magnification constant.

image8

Figure 8 The distance between the objective center and the final projected image on the tissue in response to L3 position and focal length. The red square is the selected values for the microscope.

The desired value is around 4.17cm because that seems to be the value that focuses the image at the objective working distance of 1.72cm from the objective end for the 4x and 8.5 for the 2x.

From the contours we see that increasing the focal length increases the distance. So increasing the focal length also requires moving the L3 to the right to maintain the working distance. Additionally, moving L3 too close to the dichroic may cut off some light.

The final lens picked for L3 and its position optimizes all these variables.

L4 Dynamic

The purpose of L4, the objective is to focus the light on the tissue and compensate for any optical defects in earlier stages. The final image must be at its working distance.

image9

Figure 9 The final image magnification on the tissue in response to L3 and L4 position. The red square is the selected values for the microscope.

The desired value is 0.005 - 0.01. From the contours we see that moving L3 changes the magnification, but moving L4 mostly does not.

image10

Figure 10 The distance between the objective center and the final projected image on the tissue in response to L3 and L4 position. The red square is the selected values for the microscope.

The desired value is around 4.17cm because that seems to be the value that focuses the image at the objective working distance of 1.72cm from the objective end for the 4x and 8.5 for the 2x.

From the contours we see that moving L3 changes the distance, but moving L4 mostly does not. So given a specific focal length for L3, its exact position must be fixed to get the appropriate distance.

Tissue to camera light path

For the tissue to camera path, the axis goes from the tissue towards the camera through the dichroic mirror. The center of the dichroic is the zero point, and the tissue is on the negative part of the axis and the camera is on the positive side.

Static Camera Optical Analysis

Following is the details of all the optical components in the camera path.

L4 Camera

The position of L4 (objective lens) and the tissue location are predetermined by the projector light path and the tissue location as determined above.

image11

Figure 11 The characteristics of L4 relative to the camera path.

L5

L5 is the zoom lens between the camera and the objective. Its function is to focus the image on the camera sensor plane and to zoom into the tissue to see the full projected area.

It is chosen so that it doesn’t waste too much resolution capturing tissue outside the area of interest, but also so that the camera is as close to the lens as possible.

image12

Figure 12 The characteristics of L5.

Dynamic Camera Optical Analysis

In the following, we explore how the lens values were chosen and how the lens values change the primary variables – the final position of the image and therefore camera position and its magnification. In all of this, we were looking for an ideal magnification around 3.5x (based on experimental tests). Another limitation is that the distance between the final lens (L7) and the image cannot be much less than 5cm or 6cm because the lens and camera thickness prevent that. Also, there are physical limitations on how close L5 can be to L4 because of the dichroic in the path.

L4 Dynamic camera

The position and characteristics of L4 is determined by the projector path.

L5 Dynamic

The purpose of L5 is to focus the image on the camera.

The exact placement of the lens and its focal length determines the zoom. It also controls the distance between the lens and the focused image, where the camera’s sensor needs to be placed.

image13

Figure 13 The final image magnification on the camera in response to L5 position and focal length. The red square is the selected values for the microscope.

The desired value is around 3.5. From the contours we see that increasing the focal length increases the magnification. However, moving the lens mostly keeps the magnification constant.

image14

Figure 14 The position of the final focused image (where the camera is placed) in response to L5 position and focal length. The red square is the selected values for the microscope.

The desired value is larger than about 5cm to accommodate the hardware. We also don’t want it to be too large.

From the contours we see that increasing the focal length linearly moves the focused image position. Similarly, moving the lens up, also linearly moves the focused image up.

The final lens picked and its exact position was selected to optimize the zoom and visible area of interest of the tissue.

Projector power

We measured the projector power of the blue LED as it reaches the slice through the dichroic using dichroic mirror that reflects blue to the tissue.

We varied the current using the Ceed software from 0A to 43A, the maximum. We also varied the color intensity from 1 (255) to 0.5 (127) using the parameters of the constant function in Ceed. We set the full projector area to output blue.

Following is a table of the measured power in mW at the tissue plane using a Thor PM100D with the S175C sensor. The first 4 columns was measured using the 4X, objective lens while the last two columns is from the 2X objective.

Current (A)

Intensity=1

Intensity=0.5

Intensity=0.25

*Intensity=0.25

2X (Intensity=1)

*2X (Intensity=1)

0

0.00

0.00

-0.01

-0.01

0.02

0.02

1

0.50

0.37

0.11

0.09

0.93

0.85

2

1.18

0.63

0.32

0.27

2.51

2.28

3

1.86

0.92

0.59

0.49

3.98

3.62

5

3.07

1.50

0.96

0.80

6.57

5.98

10

5.71

2.77

1.72

1.43

12.25

11.15

15

7.93

3.87

2.38

1.99

17.04

15.51

20

9.87

4.90

3.02

2.53

21.08

19.18

25

11.55

5.67

3.51

2.93

24.88

22.64

30

13.03

6.44

3.87

3.23

28.07

25.55

35

14.41

7.10

4.24

3.55

30.96

28.18

40

15.61

7.60

4.61

3.86

33.41

30.41

43

16.18

7.94

4.77

3.99

34.70

31.58

The star indicates that the column data is conditioned. The 4X 0.25 intensity and 2X data was measured on a second day than the 4X 1.0 and 0.5 intensity data. The sensor gained an offset of about 3.12 mW on the second day as measured at full 1.0 intensity at 43A. The starred columns, are adjusted proportionally to compensate.

Using the data, above 3A, the best formula for the power at a particular current, intensity, and objective is:

mW = (-0.0047*I^2 + 0.5682*I + 0.3438) * intensity * K

where

  • I is the current,

  • intensity is the intensity value between 0-1 set in a Ceed function, and

  • K is 1 for the 4X objective and 1.95 for the 2X objective.

The power at low intensity (below 0.25) was not measured, but given the projector design it’s expected to hold.

The power when the current is below about 3A did not scale according to a clear function, so it is recommended to not set the current below 3A (except to zero).